A dead-stop amperometric method based on electroactive equilibrium I-2 reve
rsible arrow 21(-) redox couple is developed for the estimation of iodine i
n iodized salts locally marketed in Pakistan Twenty iodized salt samples. p
rocured from local retailers were subjected to estimation of iodine content
by the method The iodine concentrations in the samples were found to be ve
ry divergent, ranging from as low as 4.97 mg/kg to as high as 446.38 mg/kg.
Parallel estimations. using simple thiosulphate-based titrimetric method,
were conducted and the accuracy and precision of the proposed method is dis
cussed The lower detection limit goes to close 1 mg/kg for the proposed met
hod and the present study indicated that the method is quantitatively relia
ble, as determined by its application on synthetic samples of salt containi
ng a broad range of iodine concentration, between 2.5 mg/kg to 120 mg/kg, w
ith an overall percent error in the range of 0.5 to 4.4%.