Efficiency calibration of a Ge detector in the 0.1-11.0 MeV region

Citation
S. Raman et al., Efficiency calibration of a Ge detector in the 0.1-11.0 MeV region, NUCL INST A, 454(2-3), 2000, pp. 389-402
Citations number
43
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
454
Issue
2-3
Year of publication
2000
Pages
389 - 402
Database
ISI
SICI code
0168-9002(20001111)454:2-3<389:ECOAGD>2.0.ZU;2-D
Abstract
An efficiency curve was constructed for a large-volume Ge detector located at the thermal-neutron beam line of the JRR-3M reactor at the Japan Atomic Energy Research Institute (JAERI). The calibrations for gamma -ray energies below 2754 keV were accomplished with a large number of radioactive source s and above 2754 keV with the C-12(n,gamma) and N-14(n,gamma) reactions. Th is detector and its efficiency curve were subsequently used to determine ac curately the photon emission probabilities of gamma rays from the decays of Co-56 (T-1/2 = 77 d) and Ga-66 (9.5 h) and from the Cl-35(n,gamma) reactio n. These radiation sources can serve as secondary standards for efficieney calibrations. (C) 2000 Elsevier Science B.V. All rights reserved.