Jam. Lopes et al., A large area avalanche photodiode as the VUV photosensor for gas proportional scintillation counters, NUCL INST A, 454(2-3), 2000, pp. 421-425
The feasibility of using UV-sensitive Large Area Avalanche Photodiodes (LAA
PD) as the photosensor in xenon Gas Proportional Scintillation Counters (GP
SC) is investigated. It is shown that the LAAPD can successfully replace th
e photomultiplier tube as the UV photosensor of GPSCs without degradation o
f the detector energy resolution, enabling the development of compact GPSCs
. Energy resolutions of 7.9 % and 4.4 % are obtained with a xenon-filled GP
SC for 5.9 and 22.1 keV X-rays, respectively. (C) 2000 Elsevier Science B.V
. All rights reserved.