Rationale and Objectives, Electrocardiographic (ECG) gating of cardiac magn
etic resonance (MR) imaging has been problematic for many reasons. The purp
ose of this study was to demonstrate the feasibility of using Doppler ultra
sound (VS) gating, either directly off the moving cardiac wall or the systo
lic upstroke of the arterial signal from the great vessels in neck, in alte
rnative gating modes.
Materials and Methods. A 2.5-MHz, range-gated Doppler US device was used wi
th A-mode guidance for gating directly off left ventricular wall motion. A
4- or 8.1-MHz, continuous-wave (CW) Doppler US device was used for gating o
ff the systolic upstroke from the great Vessels in the neck. The subject un
dergoing imaging held the transducer against his chest for range-gated Dopp
ler US and against his neck for 8.1-MHz CW Doppler US. The 4-MHz transducer
was strapped to the subject's neck. Modified Doppler signals were fed back
into the gating circuitry of the MR imager to achieve cardiac syn synchron
y.
Results. Cardiac gating was achieved by using both the range-gated techniqu
e directly off the cardiac wall and the CW method off blood flow from the g
reat vessels. Problems occurred with radiofrequency shielding during the ra
nge-gated method; however, these problems were almost completely removed by
use of the CW Doppler probes.
Conclusion. Doppler US gating of MR images is possible and potentially coul
d overcome many shortcomings of ECG gating. Subsequent embodiments of the t
echnique will require improved radiofrequency shielding in the range-gated
technique.