Z. Benzo et al., Influence of additives on the carryover effects of Hg during ICP-OES analysis of immunoglobulines samples, ATOM SPECT, 21(5), 2000, pp. 165-171
A method for the determination of mercury in immunoglobulines samples by in
ductively coupled plasma optical emission spectrometry (ICP-OES) is describ
ed, in which sample dilution with some additives was the only pretreatment
prior to analysis. Different chemical reagents, such as deproteinization ag
ents, and other additives, such as surfactants, were proven as potentially
useful. Residual mercury signal after samples analysis was one of our main
concern. Among the diluents studied, EDTA and Triton(R) X-100 showed more a
ccurate results. The Hg concentration in the samples, using the proposed me
thod, are in good agreement with those obtained of samples analyzed by the
official method using Cold Vapor Atomic Absorption Spectrometry (CV-AAS) wi
th partial and total destruction of the sample.