S. Kjellstrom et al., Flow immunochemical bio-recognition detection for the determination of Interleukin-10 in cell samples, J IMMUNOL M, 246(1-2), 2000, pp. 119-130
On- and off-line heterogeneous non-competitive flow immunoassays for the de
termination of Interleukin-10 are described. The sample containing IL-10 is
mixed, either on-line in a reaction coil or off-line in a test tube, with
fluorescent labelled anti-IL-10 antibodies to form an antibody-antigen comp
lex. The labelled unbound antibodies are trapped on an immobilized IL-10 co
lumn whereas the IL-10-antibody complexes are eluted and detected downstrea
m by a fluorescence detector. The optimization of the systems was performed
with respect to choice of affinity support, flow rate, carrier buffer addi
tives, pH and antibody-antigen association. Both bio recognition assays wer
e tested with a spiked cell medium and the IL-10 detection limits in this m
atrix was found to be 8 fmol using the off-line incubation mode and 40 fmol
using the on-line incubation mode. The sample through-put was 26 and 40 sa
mples per hour in the on-line and off-line incubation modes, respectively.
IL-10 identification in the sample fractions was achieved using MALDI-TOF M
S. (C) 2000 Elsevier Science B.V: All rights reserved.