It is desirable to maintain normal, conducted ventricular activation in pat
ients with dual-chamber pacemakers and preserved atrioventricular (AV) cond
uction. The shortest AV delay resulting in consistent ventricular inhibitio
n (avoiding ventricular pseudofusion) was determined by a conventional incr
emental (inside-out) technique vs the alternate decremental (outside-in) te
chnique in 20 such patients. Determinations were made in VDD mode in 20 pat
ients and DDD mode (similar to 10 beats/min faster than the intrinsic rate)
in 19. In VDD mode, the shortest AV delay avoiding ventricular pseudofusio
n was never found during inside-out testing. It was identical with both met
hods in 20 patients (50%), and shorter by 10-80 ms (mean 20 +/- 20 ms) with
the outside-in method in the remaining 10 (P = 0.004). In DDD mode, the sh
ortest AV delay resulting in consistent ventricular inhibition was found on
ly once during inside-out testing. It was the same with both methods in 13
patients (68%), and shorter by 10-20 ms (mean 14 +/- 5 ms) with the outside
-in method in the remaining 5 (26%, P = 0.18; Fisher's exact test). The sho
rtest sensed AV delay preventing ventricular pseudofusion is most likely to
be found with a decremental method (outside-in). In rare patients, it iden
tifies AV delays resulting in inhibition, while ventricular pacing persists
at longer programmable AV delays with the conventional inside-out approach
.