Performance evaluation of nine hormone assays on the Immulite 2000 (R) immunoassay system

Citation
Fl. Tello et Dm. Hernandez, Performance evaluation of nine hormone assays on the Immulite 2000 (R) immunoassay system, CLIN CH L M, 38(10), 2000, pp. 1039-1042
Citations number
7
Categorie Soggetti
Medical Research Diagnosis & Treatment
Journal title
CLINICAL CHEMISTRY AND LABORATORY MEDICINE
ISSN journal
14346621 → ACNP
Volume
38
Issue
10
Year of publication
2000
Pages
1039 - 1042
Database
ISI
SICI code
1434-6621(200010)38:10<1039:PEONHA>2.0.ZU;2-E
Abstract
We evaluated the analytical performance of the Immulite 2000(R) immunoassay analyzer (Diagnostic Products Corporation, Los Angeles, USA) based on a ne w detection technology, electrochemical luminescence. The evaluated analyte s were thyrotropin, triiodothyronine, free thyroxine. follitropin, lutropin , prolactin, cortisol, estradiol and progesterone. We tested the assay prec ision, linearity, recovery, and correlation with comparison methods for the se analytes. For most assays, within-run and between-day imprecisions were less than 8% and 10%. respectively. The linearity and recovery were accepta ble for all assays. The correlation between the Immulite 2000(R) assays and comparison methods showed satisfactory results.