Determination of gallic acid by flow injection with electrochemiluminescent detection

Citation
Yg. Sun et al., Determination of gallic acid by flow injection with electrochemiluminescent detection, ANAL LETTER, 33(15), 2000, pp. 3239-3252
Citations number
26
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL LETTERS
ISSN journal
00032719 → ACNP
Volume
33
Issue
15
Year of publication
2000
Pages
3239 - 3252
Database
ISI
SICI code
0003-2719(2000)33:15<3239:DOGABF>2.0.ZU;2-5
Abstract
Potential dependent inhibition and enhancement effects of gallic acid on th e electrochemiluminescence (ECL) of luminol in alkaline solutions were foun d. On the basis of the enhancement effect, a flow injection method was deve loped for the determination of gallic acid. The method was simple, convenie nt and sensitive. A detection limit of 1.8x10(-8) mol/L, linear range of 3. 0x10(-8) 1.0x10(-4) mol/L, relative standard deviation of 1.0% for eleven m easurements of 5.0x10(-6) mol/L of gallic acid was found. This method was s uccessfully applied to determine the content of gallic acid in Chinese gall (moshizi).