DEVELOPMENT OF A DOUBLE METAL, AC-COUPLED SILICON PAD DETECTOR

Citation
Wt. Lin et al., DEVELOPMENT OF A DOUBLE METAL, AC-COUPLED SILICON PAD DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 389(3), 1997, pp. 415-420
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
389
Issue
3
Year of publication
1997
Pages
415 - 420
Database
ISI
SICI code
0168-9002(1997)389:3<415:DOADMA>2.0.ZU;2-1
Abstract
A large area radial silicon pad detector has been produced, tested, an d installed for the WA98 experiment at CERN. This paper describes the structure and the first test results of the silicon sensors used in th is detector. Double metal technology has been used to lead the signals from the inner pads to the edge of the sensitive region. The capacita nce for each pad has been measured and is shown to vary according to t he pad size. Signals from Sr-90 radiation have been analysed, and a si gnal-to-noise ratio of over 30 has been achieved. The RMS noise varies with the capacitance of each pad, and the variation agrees with the s pecification of the preamplifier. No evidence of cross talk due to the capacitance coupling between two metal layers has been observed.