An automated electric double-sleeved vacuuming microtrephine has been devel
oped that consistently creates a 0.8 to 1.0 mm tear-free hole:in the anteri
or lens capsule. This hole is required for lens refilling to retain accommo
dation after cataract removal. The trephine is double-sleeved. The stainles
s-steel outer sleeve vacuums the internal anterior capsule; the stainless s
teel relating inner sleeve, which has a blade on the head, then slides forw
ard and cuts the taut anterior capsule. All manipulations are controlled au
tomatically by depressing a foot pedal. Using this instrument, a 0.8 to 1.0
mm tear-free microhole in the anterior capsule was created successfully in
80% of excised pig (n = 30) and eye-bank (n = 10) eyes. This trephine will
be useful for lens refilling in the future. J Cararact Refract Surg 2000;
26:1717-1721 (C) 2000 ASCRS and ESCRS.