Ni. Balalykin et al., MEASUREMENT OF THE PLASTIC SCINTILLATOR RESPONSE IN THE MAGNETIC-FIELD, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 390(3), 1997, pp. 286-292
We present the results of measurements of the light yield dependencies
for the plastic scintillators on magnetic field value and orientation
. The magnetic field value was changed from 0 up to 3.8 T. Scintillato
rs were irradiated by electrons (Cs-137 and Bi-207) and by alpha-parti
cles (Am-241). The detected light spectra in the electron irradiating
cases show two dependencies: the first one is dependence on the magnet
ic field value only and the second-light output is dependent on the ma
gnetic field orientation when the radioactive source is outside of the
scintillator. The value and behavior of the first dependence agree wi
th data published earlier. Under alpha-particle irradiation the light
output does not depend on the magnetic field value and orientation pra
ctically.