M. Darowski et al., Computer simulation of hemodynamic parameter changes by mechanical ventilation and biventricular circulatory support, METH INF M, 39(4-5), 2000, pp. 332-338
When a Bi-Ventricular Assist Device (BVAD) is used in conjunction with mech
anical ventilation (MV) of the lungs with positive intrathoracic pressure (
Pt), the latter influences hemodynamics. The aim of our study was to assess
the simultaneous influence of BVAD and MV on hemodynamics. We assumed vent
ricular pathological conditions as reduced elastances and increased rest vo
lumes. Peripheral systemic arterial resistance was assumed to have differen
t values. Data were obtained by computer simulation. Trends in main hemodyn
amic variables were compared with clinical data from literature. Simulation
showed that systemic venous, pulmonary arterial and left atrial pressures
are very sensitive to Pt (-2 to 5 mmHg).