The MOSS camera on H-1NF

Citation
Ca. Michael et al., The MOSS camera on H-1NF, REV SCI INS, 72(1), 2001, pp. 1034-1037
Citations number
9
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
1
Year of publication
2001
Part
2
Pages
1034 - 1037
Database
ISI
SICI code
0034-6748(200101)72:1<1034:TMCOH>2.0.ZU;2-V
Abstract
We have configured the modulated optical solid-state spectrometer, a recent ly developed high-resolution instrument for plasma Doppler spectroscopy, as an imaging spectroscopic camera. The camera features a wide field of view (similar to 10 degrees), large aperture (40 mm), and high spectral resoluti on nu/Delta nu greater than 10 000. The camera installation on the H-1NF He liac is described, together with the steps in the design process, including field widening. Calibration and characterization of the instrument functio n is discussed and the instrument performance is illustrated with some samp le results of spatially resolved ion temperature measurements in H-1NF. (C) 2001 American Institute of Physics.