Improvements in the evaluation of Thomson scattering data on ASDEX upgrade

Citation
B. Kurzan et al., Improvements in the evaluation of Thomson scattering data on ASDEX upgrade, REV SCI INS, 72(1), 2001, pp. 1111-1114
Citations number
2
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
1
Year of publication
2001
Part
2
Pages
1111 - 1114
Database
ISI
SICI code
0034-6748(200101)72:1<1111:IITEOT>2.0.ZU;2-5
Abstract
The avalanche diodes used as light detectors in the Thomson scattering diag nostic on the Axially Symmetric Divertor Experiment (ASDEX) Upgrade tokamak show a reduced signal-to-noise ratio during plasma discharges with high ne utron rates and during plasma edge measurements. This often makes it imposs ible to determine the electron temperature from the ratios of the scatterin g signals of the different spectral channels analyzing the Thomson scatteri ng spectrum. With the use of the maximum likelihood method combined with li nearized electron temperature characteristics the electron temperature and density are determined directly from the scattering signals without calcula ting ratios between the signals. With this method the scattering signals fo r nearly all timepoints can be evaluated and the statistical errors of the electron temperature and density are up to factors of 2 and 1.4, respective ly, smaller than achieved with the previous method. (C) 2001 American Insti tute of Physics.