A new solid sorbent passive air sampler for ketones has a silicone membrane
atop a diffusion cylindrical path length of 1.1 cm and diameter 1.3 cm abo
ve a pellet of Tenax TA coated with 10% (w/w) O-(2,3,4,5,6-pentafluorobenzy
l)hydroxylamine hydrochloride. Vapors of known concentrations approximating
their workplace permissible exposure limits of Occupational Safety and Hea
lth Administration-regulated ketones at a relative humidity (RH) of 3 +/- 1
% were generated by a syringe pump connected to a dynamic air dilution syst
em connected to an exposure chamber that allowed measurement of face veloci
ties, temperatures, exposing vapor concentrations, and RHs. The O-oxime der
ivative was desorbed with hexane, and an aliquot injected for gas chromatog
raphic analysis on a nonpolar capillary column by mass spectrometric or ele
ctron capture detection. The experimental passive sampler sampling rates in
milliliters per minute at 25 degreesC were 4.07 +/- 0.49, cyclohexanone; 6
.30 +/- 0.59, diethyl ketone; 6.31 +/- 0.31, ethyl n-butyl ketone; 3.78 +/-
0.25, methyl n-amyl ketone; 3.43 +/- 0.19, methyl n-butyl ketone; 6.48 +/-
0.64, methyl ethyl ketone; 4.37 +/- 0.43, methyl isopropyl ketone; and 4.5
7 +/- 0.17, methyl n-propyl ketone. These preliminary data show that steric
ally unhindered ketones can be sampled by the passive sampler as well as al
dehydes.