Trace level determination of lead by flame AAS using on-line separation and preconcentration

Citation
Mc. Yebra-biurrun et al., Trace level determination of lead by flame AAS using on-line separation and preconcentration, ATOM SPECT, 21(6), 2000, pp. 225-228
Citations number
8
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ATOMIC SPECTROSCOPY
ISSN journal
01955373 → ACNP
Volume
21
Issue
6
Year of publication
2000
Pages
225 - 228
Database
ISI
SICI code
0195-5373(200011/12)21:6<225:TLDOLB>2.0.ZU;2-R
Abstract
A flow injection (FI) on-line preconcentration method was developed for the flame atomic absorption spectrometry (FAAS) determination of trace amounts of lead in alloy samples. Lead was preconcentrated on a minicolumn packed with a poly(aminophosphonic acid) chelating resin (PAPhA), and subsequently eluted using a small volume of hydrochloric acid (2 mol/L) directly into t he nebulizer burner system of the spectrometer. The procedure resulted in a n enhancement factor of 14. A detection limit of 6.3 mug/L (3 sigma) was ob tained using a time-based technique with 30 s preconcentration; a 2.1% RSD was readily achieved for 11 measurements of 50 mug/L Pb. The sample through put was about 80/h. The procedure was applied to the determination of lead in alloy samples with high tolerance to interferences,and the results obtai ned for standard reference materials were in good agree ment with the certi fied values.