Mc. Yebra-biurrun et al., Trace level determination of lead by flame AAS using on-line separation and preconcentration, ATOM SPECT, 21(6), 2000, pp. 225-228
A flow injection (FI) on-line preconcentration method was developed for the
flame atomic absorption spectrometry (FAAS) determination of trace amounts
of lead in alloy samples. Lead was preconcentrated on a minicolumn packed
with a poly(aminophosphonic acid) chelating resin (PAPhA), and subsequently
eluted using a small volume of hydrochloric acid (2 mol/L) directly into t
he nebulizer burner system of the spectrometer. The procedure resulted in a
n enhancement factor of 14. A detection limit of 6.3 mug/L (3 sigma) was ob
tained using a time-based technique with 30 s preconcentration; a 2.1% RSD
was readily achieved for 11 measurements of 50 mug/L Pb. The sample through
put was about 80/h. The procedure was applied to the determination of lead
in alloy samples with high tolerance to interferences,and the results obtai
ned for standard reference materials were in good agree ment with the certi
fied values.