Flow-through analysis of Al in seawater by fluorometric detection with theuse of Lumogallion

Citation
H. Obata et al., Flow-through analysis of Al in seawater by fluorometric detection with theuse of Lumogallion, FIELD A C T, 4(6), 2000, pp. 274-282
Citations number
22
Categorie Soggetti
Environment/Ecology,"Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
FIELD ANALYTICAL CHEMISTRY AND TECHNOLOGY
ISSN journal
1086900X → ACNP
Volume
4
Issue
6
Year of publication
2000
Pages
274 - 282
Database
ISI
SICI code
1086-900X(2000)4:6<274:FAOAIS>2.0.ZU;2-F
Abstract
A flow-through analytical method has been developed for the determination o f trace amounts of aluminum in seawater with the use of Lumogallion fluorom etric detection. Because iron was selectively removed by passing the sample through an 8-quinolinol immobilized chelating resin column (MAF-8HQ), this method could be applied to samples collected at the hydrothermal oceanic r egime. Several known organic complexing agents were studied as a model grou p to examine the effect of naturally occurring organic ligands on the detec tion of aluminum in seawater. The detection limit was 0.17 nM for 10 ml of sample. The standard deviation was 2.7% at 2 nM of aluminum. This method pe rmitted the sensitive, precise, and rapid determination of subnanomolar Al in seawater without contamination onboard ship. (C) 2000 John Wiley & Sons, Inc.