Photon spectra are reported that result from the scatter of diagnostic
X rays at an appropriate water phantom that represents a patient. The
tube voltages considered are between 52 kV and 110 kV, the scatter an
gles range from 10 degrees to 142 degrees to the normal radiation inci
dence direction. All spectral measurements were performed with a high-
purity germanium detector. Spectral photon fluences are computed from
the measured pulse height distribution by using an unfolding procedure
. The required response functions of the detection system were obtaine
d by using Monte Carlo methods. Reference is made to a catalogue compi
ling 35 spectra of scattered X rays in diagnostics resulting from a hu
man substitute, together with information on relevant primary field pa
rameters as well as air kerma for the scattered X rays.