Dg. Mascher et al., High-performance liquid chromatographic-tandem mass spectrometric determination of 3-hydroxypropylmercapturic acid in human urine, J CHROMAT B, 750(1), 2001, pp. 163-169
A sensitive and specific high-performance liquid chromatographic-tandem mas
s spectrometric (HPLC-MS-MS) method was developed for the determination of
3-hydroxypropylmercapturic acid (3-HPMA) in human urine. Samples were extra
cted using ENV+ cartridges and then injected onto a C8 Superspher Select B
column with acetonitrile and formic acid as eluent (5:95, v/v). N-Acetylcys
teine was used as internal standard for HPLC-MS-MS. Linearity was given in
the tested range of 50-5000 ng/ml urine. The limit of quantification was 50
ng/ml. Precision, as C.V., in the tested range of 50-5000 ng/ml was 1.47-6
.04%. Accuracy ranged from 87 to 114%. 3-HPMA was stable in human urine at
37 degreesC for 24 h. The method was able to quantify 3-HPMA. in urine of n
on-smokers and smokers. (C) 2001 Elsevier Science B.V. All rights reserved.