HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF TAURINE AND HYPOTAURINE USING 3,5-DINITROBENZOYL CHLORIDE AS DERIVATIZING REAGENT

Citation
N. Masuoka et al., HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF TAURINE AND HYPOTAURINE USING 3,5-DINITROBENZOYL CHLORIDE AS DERIVATIZING REAGENT, Journal of chromatography B. Biomedical applications, 660(1), 1994, pp. 31-35
Citations number
21
Categorie Soggetti
Chemistry Analytical
Journal title
Journal of chromatography B. Biomedical applications
ISSN journal
15726495 → ACNP
Volume
660
Issue
1
Year of publication
1994
Pages
31 - 35
Database
ISI
SICI code
Abstract
A method for the determination of taurine and hypotaurine in biologica l samples involving the preparation of their 3,5-dinitrobenzoyl deriva tives followed by HPLC was established. Taurine and hypotaurine in aqu eous media were reacted with 3,5-dinitrobenzoyl chloride in the presen ce of triethylamine to prepare 3,5-dinitrobenzoyl derivatives. These d erivatives were separated on a C-18 reversed-phase column and detected by recording the absorbance at 254 nm. Derivatives of taurine and hyp otaurine were obtained in yields of 91.4 +/- 3.3 and 85.6 +/- 2.6%, re spectively. The calibration graphs for taurine and hypotaurine were li near between 2.5 and 500 mu M with correlation coefficients of 0.999. The method was applied to the determination of taurine and hypotaurine in human and rat urine and blood and in rat liver and heart.