Gp. Suarez-fernandez et al., Analysis of major, minor and trace elements in coal by radioisotope X-ray fluorescence spectrometry, FUEL, 80(2), 2001, pp. 255-261
The evaluation of the energy dispersive X-ray fluorescence spectroscopy met
hod, for the analysis of major, minor and trace elements present in the min
eral matter of coal, using an apparatus with sources of radioisotope elemen
ts, is described. Analyses were performed on both coal and coal ash samples
and the results were compared. From these results it can be concluded that
, the precision and accuracy obtained is lower than that of those XRF techn
iques in which an X-ray tube is used as an X-ray source. However, it is sat
isfactory for the rapid routine analysis of some major and minor elements.
The advantage of the method is its potentially wide application in industry
for quick analysis. (C) 2000 Elsevier Science Ltd. All rights reserved.