This report describes new treadmill ergometer designed to measure the verti
cal and horizontal ground reaction forces produced by the left and right le
gs during walking. It was validated by static and dynamic tests. Non linear
ity was from 0.2% (left vertical force) to 1.4% (right antero-posterior for
ce). The resonance frequency was from 219 (right vertical direction) to 58
Hz (left medio-lateral direction). A calibration "leg", an air jack in seri
es with a strain gauge, was developed and used to produce force signals com
parable to those obtained during human locomotion. The mean differences bet
ween the force measured by the calibration leg and treadmill ergometer at 5
km h(-1) were 3.7 N (0.7%) for the left side and 6.5 N (1.2%) for the righ
t. Measurements obtained during human walking showed that the treadmill erg
ometer has considerable potential for analysing human gait. (C) 2000 Elsevi
er Science Ltd. All rights reserved.