R. Andreoli et al., HPLC ANALYSIS OF FAT-SOLUBLE VITAMINS ON STANDARD AND NARROW-BORE COLUMNS WITH UV, ELECTROCHEMICAL AND PARTICLE-BEAM MS DETECTION, Chromatographia, 44(11-12), 1997, pp. 605-612
Citations number
15
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
The application of different detection techniques; UV, electrochemical
(ED) and particle beam-mass spectrometry (PB-MS) for the high-perform
ance liquid chromatography analysis of fat-soluble vitamins were studi
ed. Separation of all compounds examined was performed using both C8 s
tandard-bore and narrow-bore columns for UV and MS detection, whereas
in the case of HPLC-ED, a C8 standard-bore column only was used. Detec
tor responses were studied in terms of linear range, detection limit a
nd precision. The HPLC methods with UV, PB-MS and electrochemical dete
ction were applied to the assay of vitamins A and E in an infant-formu
la, powdered-milk sample. Recovery studies showed good results both fo
r vitamins A and E, within-run precision being 2-4 % for LC-UV and LC-
ED techniques. In addition, excellent agreement between the results ob
tained with the detection systems explored was found.