European and international standards lay down criteria for the size-selecti
ve aerosol sampling in occupational hygiene, Aerosol samplers are supposed
to match these target sampling criteria, This study focused on 15 aerosol s
amplers used to sample the conventional respirable fraction, An aerodynamic
particle sizer (APS) method was used to measure the sampling efficiency of
the samplers in a low-velocity wind tunnel, Polydisperse coat dust was gen
erated as the test aerosol, The data were fitted by an appropriate mathemat
ical model, For some instruments the results show serious deviations from t
he conventional target curve, whereas other devices meet the convention qui
te well, The Row rate of certain cyclone-separator-based instruments was op
timized to adjust their sampling efficiency, The mass concentration bias an
d accuracy of the samplers were calculated for a number of ranges of partic
le size distributions of aerosols commonly found in industrial workplaces,
Finally, the performance of each sampler was evaluated using bias and accur
acy maps, Most of these samplers are suitable for sampling the CEN-ISO-ACGI
H respirable fraction of aerosols, but several require modification of the
flow rate, For real industrial situations, the rough knowledge of the aeros
ol size distribution can guide the choice of an appropriate sampling techni
que, (C) 2001 British Occupational Hygiene Society, Published by Elsevier S
cience Ltd, All rights reserved.