Mr. Feddah et al., In-vitro characterisation of metered dose inhaler versus dry powder inhaler glucocorticoid products: Influence of inspiratory flow rates, J PHARM P S, 3(3), 2000, pp. 317-324
Purpose: To study the influence of inspiratory flow rate on the fine partic
le mass and the particle size distribution for metered dose inhaler (MDI) a
nd dry powder inhaler (DPI) glucocorticoid products, in vitro. To compare t
he performance of MDI and DPI inhalers containing the same drug and strengt
h at an impaction flow rate of 60 L/min. Methods: The Marple Miller cascade
impactor model 150 and 160 were used to characterise several glucocorticoi
d MDI and DPI products at different simulated inspiratory flow rates (30 L/
min, 60 L/min and 90 L/min). Following the actuation of one single inhaler
puff the amount of drug deposited in each stage of the impactor was quantif
ied using high performance liquid chromatography with UV detection at 242 n
m. The size distribution of the primary particles of DPI products was measu
red by laser diffraction. Results: DPIs were significantly more dependent o
n impaction flow rate than MDIs. Except for Pulmicort(R), the fine particle
mass FPM delivered from the MDI products was significantly higher than tha
t delivered from the DPI aerosols. Conclusions: Although the metered dose i
nhaler is the older technology it exhibits greater respirable dose in vitro
than newer dry powder inhaler devices. Care should be taken when shifting
from one inhaler dosage form to another because this may affect the actual
dose delivered to the lung. Further in vivo studies may be warranted in lig
ht of these findings.