RESPIROMETER FOR SMALL SAMPLES WITH AUTOMATIC REGISTRATION

Authors
Citation
F. Floto, RESPIROMETER FOR SMALL SAMPLES WITH AUTOMATIC REGISTRATION, Respiration physiology, 108(3), 1997, pp. 247-252
Citations number
12
Categorie Soggetti
Respiratory System",Physiology
Journal title
ISSN journal
00345687
Volume
108
Issue
3
Year of publication
1997
Pages
247 - 252
Database
ISI
SICI code
0034-5687(1997)108:3<247:RFSSWA>2.0.ZU;2-A
Abstract
The respiration chamber containing the sample is a small glass test tu be with a closure of KOH solution. During registration the gas volume is kept constant by adjusting the pressure through a feedback mechanis m: a microscope projects the picture of the meniscus of KOH onto a pho todiode connected to a pen recorder with a vertically moving pen. A le velling bulb with water mounted on the pen transmits the pressure adju stments to the KOH solution through a water filled tubing. The accurac y is +/-0.1 mm water column (1 Pa) and the minimum air volume is 10 nl , so that an oxygen consumption down to 1 pl can be seen. (C) 1997 Els evier Science B.V.