An algorithm for stereotactic localization by computed tomography or magnetic resonance imaging
Citation
Jr. Dai et al., An algorithm for stereotactic localization by computed tomography or magnetic resonance imaging, PHYS MED BI, 46(1), 2001, pp. N1-N7
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
SICI code
0031-9155(200101)46:1<N1:AAFSLB>2.0.ZU;2-L
Abstract
Stereotactic localization of an intracranial lesion by computed tomography
or magnetic resonance imaging requires the use of a head frame that is fixe
d to the skull of the patient. To such head frames are attached either N-sh
aped or V-shaped localization rods. Because of patient positioning, the tra
nsverse imaging slices may not be parallel to the frame base; a coordinate
transformation algorithm that takes this possibility into consideration is
crucial. Here we propose such an algorithm for a head frame with V-shaped l
ocalization rods. Our algorithm determines the transformation matrix betwee
n the image coordinate system of a transverse image and the frame coordinat
e system. The determining procedure has three steps: (a) calculation of the
oblique angles of a transverse image relative to the head frame and calcul
ation of the image magnification factor; (b) determination of the coordinat
es of four central markers in both coordinate systems; and (c) determinatio
n of the 3 x 3 transformation matrix by using the coordinates of the four m
arkers. This algorithm is robust in principle and is useful for improving t
he accuracy of localization.