Apja. Markus et al., New technique for diagnosis and monitoring of alcaptonuria: quantificationof homogentisic acid in urine with mid-infrared spectrometry, ANALYT CHIM, 429(2), 2001, pp. 287-292
For the first time we used Fourier transform-infra-red (FT-IR) spectrometry
in the investigation of an inborn error of metabolism, i.e. alcaptonuria,
which expresses itself by the excretion of high concentrations of homogenti
sic acid in urine. The 200 homogentisic acid-spiked urine samples, ranging
from 0 to 60 mmol l(-1), were analysed in the mid-IR region using an attenu
ated total reflection accessory. Quantification is performed with partial l
east squares regression analysis. The standard errors of calibration, cross
validation and prediction in the IR-predicted homogentisic acid concentrat
ion are 0.64, 0.78 and 0.79 mmol l(-1),respectively, with R-2 > 0.998. The
2 sigma detection limit is 2.0 mmol l(-1) and the intra- and inter-assay co
efficients of variation are below 5% for three concentration levels. The cl
ear difference in the concentrations of homogentisic acid in the urine of f
our alcaptonuria patients compared to seven control samples demonstrates th
e potential suitability of IR analysis for the diagnosis and monitoring of
alcaptonuria patients. (C) 2001 Elsevier Science B.V. All rights reserved.