High performance liquid chromatography with fluorimetric detection for thedetermination of total homocysteine in human plasma: method and clinical applications
St. Chou et al., High performance liquid chromatography with fluorimetric detection for thedetermination of total homocysteine in human plasma: method and clinical applications, ANALYT CHIM, 429(2), 2001, pp. 331-336
Homocysteine (Hcy), an amino acid derived from methionine is an independent
risk factor for coronary artery disease. We have developed a convenient hi
gh performance liquid chromatography assay with fluorimetric detection for
the accurate determination of Hey in human plasma. This assay involves dire
ct addition of human plasma to methanolic monobromobimane for simultaneous
protein precipitation and thiol derivatization. The assay was validated by
adding authentic homocysteine and homocystine to plasma samples with recove
ry rates between 95.5 and 102.4%. The relative standard deviations for intr
aday (n = 7) and interday (n = 6, 10 consecutive days) analysis were 0.7-3.
9 and 0.7-4.0%, respectively. Reference values of plasma total homocysteine
(tHcy) have been determined for men and women, showing a significant diffe
rence (P < 0.05) between genders. Preliminary results for occluded coronary
artery disease and for maintenance hemodialysis are presented. (C) 2001 El
sevier Science B.V. All rights reserved.