A low-loss, ultrahigh vacuum compatible helium cryostat without liquid nitrogen shield

Citation
H. Okamoto et Dm. Chen, A low-loss, ultrahigh vacuum compatible helium cryostat without liquid nitrogen shield, REV SCI INS, 72(2), 2001, pp. 1510-1513
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
2
Year of publication
2001
Pages
1510 - 1513
Database
ISI
SICI code
0034-6748(200102)72:2<1510:ALUVCH>2.0.ZU;2-U
Abstract
An ultrahigh vacuum compatible compact helium cryostat with a 48 h hold tim e for a 2.5 l of liquid helium capacity has been realized. The design explo its the large enthalpy of the gas phase helium, by using two layers of vapo r-cooled radiation shield in place of the conventional liquid nitrogen jack et. The mechanical rigidity, the absence of liquid nitrogen bubbling, and a low helium boil-off rate make this cryostat particularly suitable for appl ications that require a low vibration environment such as scanning probe mi croscopy. (C) 2001 American Institute of Physics.