Y. Lu et al., Thermal chondroplasty with radiofrequency energy - An in vitro comparison of bipolar and monopolar radiofrequency devices, AM J SP MED, 29(1), 2001, pp. 42-49
The purpose of this study was to examine the in vitro effects of three radi
ofrequency energy devices (two bipolar devices and one monopolar device) fo
r the performance of thermal chondroplasty. Thirty-two fresh bovine femoral
osteochondral sections (approximately 3 x 4 x 5 cm) from eight cows were d
ivided into four groups (three treatment patterns and one sham-operated gro
up with eight specimens per group). The three treatment patterns consisted
of 1) radiofrequency energy delivered by a mechanical jig at 1 mm/sec in a
contact mode (50 g of pressure), 2) radiofrequency energy delivered by a me
chanical jig at 1 mm/sec in a noncontact mode (1 mm between probe tip and a
rticular cartilage surface), and 3) radiofrequency energy smoothing of abra
ded cartilage during arthroscopic visualization. Thermal smoothing of the a
braded cartilage surface was accomplished with all three devices. Significa
nt chondrocyte death, as determined by confocal laser microscopy and cell v
iability staining, was observed with each device. The bipolar radiofrequenc
y systems penetrated 78% to 92% deeper than the monopolar system. The bipol
ar systems penetrated to the level of the subchondral bone in ail osteochon
dral sections during arthroscopically guided paintbrush pattern treatment.
Radiofrequency energy should not be used for thermal chondroplasty until fu
rther work can establish consistent methods for limiting the depth of chond
rocyte death while still achieving a smooth articular surface.