T. Storm et al., Exact mass measurements on-line with high-performance liquid chromatography on a quadrupole pass spectrometer, ANALYT CHEM, 73(3), 2001, pp. 589-595
Exact mass measurements were performed on-line with high-performance liquid
chromatography on a quadrupole mass spectrometer. Compounds with molecular
weights from 98 to 797, mainly aromatic sulfonates and sulfonamides, were
analyzed with electrospray ionization in positive or negative mode. Interna
l mass calibration compounds were continuously added after separation. A Ga
ussian fit of the mass errors of 808 individual measurements (concentration
s of 1-10 mg/L, 20-200 ng absolute on column) resulted in a mean error of 0
.1 mmu (0.45 ppm) and a standard deviation sigma of 1.5 mmu (5.4 ppm). The
99.7% confidence intervals (3 sigma) were +/-4.5 mmu (+/-16.2 ppm) for sing
le mass measurements. Averaging 10 measurements further reduced the errors
to less than +/-1.5 mmu (+/-5 ppm). Isobaric interferences with ions result
ing from the mass calibrants were avoided by the use of complementary mass
calibrants. The results were verified (differences below +/-4.5 mmu) with a
LC/oa-TOFMS. Limited mass range chromatograms were used to enhance selecti
vity in the analysis of mixtures. The method was applied to determine the e
lemental composition of a potential dye metabolite detected in anaerobicall
y treated textile wastewater.