M. Ehlers et al., Design and set-up of an external ion source coupled to a quadrupole-ion-trap reflectron-time-of-flight hybrid instrument, EUR J MASS, 6(4), 2000, pp. 377-385
Further results of a new set-up for storing ions generated in an external i
on source and trapped in a quadrupole ion trap in combination with a reflec
tron time-of-flight mass spectrometer (ReTOF) are presented. As a result of
this set-up, jet-cooled molecules are introduced as neutrals into the exte
rnal ion source and ionized by multiphoton absorption processes. A selected
mass range of the generated ions can be accelerated into the ion trap to b
e stored without any buffer gas, The ions are not dispersed for a time dura
tion of several milliseconds in the ion trap, but remain coherent in space
and kinetic energy and move in narrow ion clouds describing Lissajous traje
ctories. The advantage of the external ion source and its influences on the
parameters of both the external ion source and the ion trap will be discus
sed with respect to the signals obtained and compared with SIMION calculati
ons.