Citation
Lc. Levy et al., Spinning drop tensiometry using a square section sample tube, J COLL I SC, 234(2), 2001, pp. 442-444
Citations number
13
Categorie Soggetti
Physical Chemistry/Chemical Physics
Journal title
JOURNAL OF COLLOID AND INTERFACE SCIENCE
ISSN journal
00219797 → ACNP
Volume
234
Issue
2
Year of publication
2001
Pages
442 - 444
Database
ISI
SICI code
0021-9797(20010215)234:2<442:SDTUAS>2.0.ZU;2-1
Abstract
When measuring interfacial tensions, conventional spinning drop tensiometry requires knowledge of the index of refraction of the denser, outer fluid p hase, in order to correct for the magnification of the inner, spinning drop . We have modified a spinning drop tensiometer to accommodate a square sect ion sample tube, which does not require this correction. Tests performed us ing calibration rods are used to demonstrate this fact. The interfacial ten sion of toluene and water is also measured using this new design. The ensui ng results are in agreement with measurements obtained using a conventional circular sample tube. (C) 2001 Academic Press.