Comparative testing of three different coatings for the immobilization of t
he anti-Salmonella enteritidis antibody onto piezoelectric (PZ) crystals wi
th gold, silver and palladium electrodes was done. In terms of stability, s
implicity and magnitude of the frequency change caused by antibody binding
to the coating, polyethyleneimine film gave the better results for the immo
bilization of antibody as compared with the (gamma -aminopropyl)trimethoxys
ilane film or the self-assembled monolayer of 3-mercaptopropionic acid. To
reduce the production cost, the small-sized crystal with silver electrodes
was used to fabricate the disposable antibody-modified PZ transducer. The r
apid detection of Salmonella enteritidis was successfully performed using t
he antibody-modified PZ immunosensor. A detection limit of 1 x 10(5) cells/
ml and an assay time of 35 min were achieved. (C) 2001 Elsevier Science B.V
. All rights reserved.