The dose mapping system for the electromagnetic calorimeter of the BaBar experiment at SLAC

Citation
B. Camanzi et al., The dose mapping system for the electromagnetic calorimeter of the BaBar experiment at SLAC, NUCL INST A, 457(3), 2001, pp. 476-486
Citations number
19
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
457
Issue
3
Year of publication
2001
Pages
476 - 486
Database
ISI
SICI code
0168-9002(20010121)457:3<476:TDMSFT>2.0.ZU;2-7
Abstract
An array of 116 p-channel radiation-sensitive MOSFET transistors (RadFETs) has been commissioned for the BaBar experiment at the PEP-II accelerator at the Stanford Linear Accelerator Center (SLAC). The aim of this system is t o map the integrated dose absorbed by different regions of the calorimeter during the course of the experiment. This paper reports the design, charact erization and installation of the sensors and briefly describes the readout system designed by co-workers. The response at low doses and the temperatu re coefficient for 130RadFETs has been determined. At a "read" current of 1 60 muA an average shift in voltage of 51 +/- 4 mV was measured at an integr ated dose of 30 rad with a temperature coefficient of 2.3 +/- 0.3 mV/degree sC. Calibration for high doses is also described. The system provides a nov el method for real-time measurements of the spatial distribution of ionizin g radiation dose within a large particle physics experiment. (C) 2001 Elsev ier Science B.V. All rights reserved.