The Umbella high-frequency EPR facility

Citation
Pjm. Van Bentum et al., The Umbella high-frequency EPR facility, APPL MAGN R, 19(3-4), 2000, pp. 479-484
Citations number
3
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
APPLIED MAGNETIC RESONANCE
ISSN journal
09379347 → ACNP
Volume
19
Issue
3-4
Year of publication
2000
Pages
479 - 484
Database
ISI
SICI code
0937-9347(2000)19:3-4<479:TUHEF>2.0.ZU;2-#
Abstract
The Umbella high-Frequency electron paramagnetic resonance (EPR) facility i s a part of the Nijmegen High Field Magnet Laboratory with spacial emphasis on multifrequency EPR For chemical and biological applications. At present the facility has various solid-state sources available for the frequency r ange between 95 and 400 GHz, combined with magnetic fields up to 30 T. For frequencies above 400 GHz a Far-infrared laser is used. With superheterodyn e detection techniques a typical dynamical range of 130 dB is obtained, wit h single-mode resonators, oversized cylindrical and Fabry-Perot resonators. Some examples will be shown to demonstrate the power of multifrequency EPR in high-spin systems.