Easily adjustable pulse shape discrimination (PSD) system effectively opera
ting in the wide range of registered energies is offered. Shape distortions
of the detector output signal due to charge losses are clearly visible if
one carries out shape analysis of the spectrometric signal for pulses resto
red up to the rectangular shape. The essence of the method consists in the
analysis of difference of a restored signal form from rectangular. For this
purpose the area of a pulse is measured within the given interval at once
after incoming pulse and within the same time interval delayed concerning f
irst. In case of an inequality of these areas, the signal is considered def
ormed and is not exposed to further analysis. Schematic decision and operat
ion mode of PSD system is analysed, pulses diagrams are presented. The deve
loped PSD system was applied in gamma-spectrometer with planar geometry CdT
e detector with dimensions 5 x 5 x 3 mm(3). Detector had m-s-m structure wi
th gold electrode. Initial energy resolution of the detector was 17keV at 6
62keV and peak-to-Compton ratio equal to 1.8. The PSD application improves
energy resolution up to 3.2 keV and peak-to-Compton ratio to about 5.0. At
this essential reduction of count rate (up to 75%) in a maximum of a total
absorption peak is observed. But the energy resolution of about 6 keV can b
e received at insignificant (not more than 20%) reduction of count rate. Sp
ectra of Cs-137 and Ba-133 obtained with and without PSD system application
are shown. Diagrams of energy resolution at 662 keV line as well as count
rate in the total absorption peak maximum versus pulse shape discrimination
level are presented. (C) 2001 Elsevier Science B.V. All rights reserved.