A simple calorimeter for fast adiabatic heat capacity measurements from 15to 300 K based on closed cycle cryocooler

Citation
I. Catarino et G. Bonfait, A simple calorimeter for fast adiabatic heat capacity measurements from 15to 300 K based on closed cycle cryocooler, CRYOGENICS, 40(7), 2000, pp. 425-430
Citations number
9
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Mechanical Engineering
Journal title
CRYOGENICS
ISSN journal
00112275 → ACNP
Volume
40
Issue
7
Year of publication
2000
Pages
425 - 430
Database
ISI
SICI code
0011-2275(200007)40:7<425:ASCFFA>2.0.ZU;2-X
Abstract
A simple, cryogen-free and inexpensive experimental setup for fast heat cap acity measurements of solids from 15 to 300 K is presented. It consists of a thermally controlled cell, coupled to the cold finger of a Gifford-Mac Ma hon cryocooler, containing two cheap Pt thin film resistors as thermometers : one is simultaneously the sample holder, the sample heater and the sample thermometer; the other resistor is used for temperature control. This calo rimeter allows adiabatic specific heat measurements in the whole temperatur e range in less than one hour. The heat capacity results for a 106 mg coppe r sample match the tabulated values within 2% for T>20 K. This system was u sed to measure the specific heat of UFexAl12-x with sample masses as low as 26 mg without performance degradation. (C) 2001 Elsevier Science Ltd. All rights reserved.