P. Mattila et al., Determination of flavonoids in plant material by HPLC with diode-array andelectro-array detections, J AGR FOOD, 48(12), 2000, pp. 5834-5841
A high-performance liquid chromatographic (HPLC) method with in-line connec
ted diode-array (DAD) and electro-array (EC) detection to identify and quan
tify 17 flavonoids in plant-derived foods is described. Catechins were extr
acted from the samples using ethyl acetate, and quantification of these com
pounds was performed with the EC detector. Other flavonoids were quantified
with DAD after acid hydrolysis. The methods developed were effective for t
he determination of catechins and other flavonoids in plant-derived foods.
Responses of the detection systems were linear within the range evaluated,
20-200 ng/injection (DAD) and 20-100 ng/injection (EC), with correlation co
efficients exceeding 0.999. Coefficient of variation was under 10.5%, and r
ecoveries of flavonoids ranged from 70 to 124%. Purity of the flavonoid pea
ks was confirmed by combining the spectral and voltammetric data.