On a single-counts rate stabilization scheme employed in the imaging camera of the TACTIC gamma-ray telescope

Citation
N. Bhatt et al., On a single-counts rate stabilization scheme employed in the imaging camera of the TACTIC gamma-ray telescope, MEAS SCI T, 12(2), 2001, pp. 167-171
Citations number
9
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
MEASUREMENT SCIENCE & TECHNOLOGY
ISSN journal
09570233 → ACNP
Volume
12
Issue
2
Year of publication
2001
Pages
167 - 171
Database
ISI
SICI code
0957-0233(200102)12:2<167:OASRSS>2.0.ZU;2-C
Abstract
A cost-effective methodology for operating the 349-pixel Cerenkov light ima ging camera of the imaging clement of the TACTIC gamma -ray telescope array at stable single-counts rate and safe anode current values has successfull y been tried out despite variations in the light of the night sky experienc ed by these pixels from time to time. Important details of this unconventio nal scheme are discussed and reference is made to the methods evolved for i n situ absolute and relative gain calibrations of the camera pixels, which air particularly required for this mode of operation.