P. Karaiskos et al., Dosimetry of Ir-192 wires for LDR interstitial brachytherapy following theAAPM TG-43 dosimetric formalism, MED PHYS, 28(2), 2001, pp. 156-166
Citations number
22
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Implementation of the AAPM Task Group 43 dosimetric formalism for Ir-192 wi
res used as interstitial sources in low dose-rate (LDR) brachytherapy appli
cations is investigated. Geometry factors, dose-rate constant values, radia
l dose functions, and anisotropy functions, utilized in this formalism, wer
e calculated for various lengths of all commercially available wire source
designs by means of a well-established Monte Carlo simulation code and an i
mproved modification of the Sievert integral method. Results are presented
in the form of look up tables that allow interpolation for dose-rate calcul
ations around all practically used wire lengths, with accuracy acceptable f
or clinical applications. (C) 2001 American Association of Physicists in Me
dicine.