One of the important features of the Siemens Virtual Wedge (VW) is that the
VW factor (VWF) is approximately equal to unity for all beams with a total
deviation for a given wedge no greater than 0.05, as specified by Siemens.
In this note we report the observed dependence of VWF on dose calibration
(cGy/MU), monitor units (MU), and beam tuning for a Primus, a linear accele
rator with two dose-rate ranges available for VW operation. The VWF is defi
ned as the ratio of doses measured on the beam central axis for the wedge f
ield to the open field; the open field dose is always measured with the nom
inal high dose-rate beam. When VW operates in the high dose-rate range, the
VWF is independent of calibration (cGy/MU). When VW works in the low dose-
rate range, the VWF varies linearly with the calibration of the low dose-ra
te mode. For a linear accelerator that has only one dose-rate range for VW,
there is no observable dependence of VWF on the calibration. We also studi
ed the monitor unit dependence of VWF. A discontinuity in VWF was observed
at the switching point between the high and low dose-rate ranges. Working w
ith Siemens, we have investigated causes of this discontinuity. As a result
of this investigation, the discontinuity in VWF as a function monitor unit
is practically removed. (C) 2001 American Association of Physicists in Med
icine.