Measurement of the dose deposition characteristics of x-ray fluoroscopy beams in water

Citation
Ka. Fetterly et al., Measurement of the dose deposition characteristics of x-ray fluoroscopy beams in water, MED PHYS, 28(2), 2001, pp. 205-209
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
28
Issue
2
Year of publication
2001
Pages
205 - 209
Database
ISI
SICI code
0094-2405(200102)28:2<205:MOTDDC>2.0.ZU;2-6
Abstract
The purpose of this study was to characterize the x-ray dose distribution o f fluoroscopy beams by measuring their percent depth dose curves and latera l dose profiles in a water phantom. Percent depth dose curves were measured near the surface with an Attix parallel plate chamber and deep within the water phantom with a Farmer-type cylindrical chamber. Percent depth dose cu rves were compared to published data where applicable. Lateral dose profile s were measured at depths of 2, 5, 10, and 15 cm in phantom with a Farmer c hamber. Pulsed, 50 mA x-ray beams with peak tube potentials of 60, 80, 100, and 120 kV and half value layers of 1.89, 2.52, 3.20, and 4.09 mm Al, resp ectively, were investigated. (C) 2001 American Association of Physicists in Medicine.