Jh. Siewerdsen et Da. Jaffray, Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter, MED PHYS, 28(2), 2001, pp. 220-231
Citations number
37
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
A system for cone-beam computed tomography (CBCT) based on a flat-panel ima
ger (FPI) is used to examine the magnitude and effects of x-ray scatter in
FPI-CBCT volume reconstructions. The system is being developed for applicat
ion in image-guided therapies and has previously demonstrated spatial resol
ution and soft-tissue visibility comparable or superior to a conventional C
T scanner under conditions of low x-ray scatter. For larger objects consist
ent with imaging of human anatomy (e.g., the pelvis) and for increased cone
angle (i.e., larger volumetric reconstructions), however, the effects of x
-ray scatter become significant. The magnitude of x-ray scatter with which
the FPI-CBCT system must contend is quantified in terms of the scatter-to-p
rimary energy fluence ratio (SPR) and scatter intensity profiles in the det
ector plane, each measured as a function of object size and cone angle. For
large objects and cone angles (e.g., a pelvis imaged with a cone angle of
6 degrees), SPR in excess of 100% is observed. Associated with such levels
of x-ray scatter are cup and streak artifacts as well as reduced accuracy i
n reconstruction values, quantified herein across a range of SPR consistent
with the clinical setting. The effect of x-ray scatter on the contrast, no
ise, and contrast-to-noise ratio (CNR) in FPI-CBCT reconstructions was meas
ured as a function of SPR and compared to predictions of a simple analytica
l model. The results quantify the degree to which elevated SPR degrades the
CNR. For example, FPI-CBCT images of a breast-equivalent insert in water w
ere degraded in CNR by nearly a factor of 2 for SPR ranging from similar to
2% to 120%. The analytical model for CNR provides a quantitative understand
ing of the relationship between CNR, dose, and spatial resolution and allow
s knowledgeable selection of the acquisition and reconstruction parameters
that, for a given SPR, are required to restore the CNR to values achieved u
nder conditions of low x-ray scatter. For example, for SPR=100%, the CNR in
FPI-CBCT images can be fully restored by: (1) increasing the dose by a fac
tor of 4 (at full spatial resolution); (2) increasing dose and slice thickn
ess by a factor of 2; or (3) increasing slice thickness by a factor of 4 (w
ith no increase in dose). Other reconstruction parameters, such as transaxi
al resolution length and reconstruction filter, can be similarly adjusted t
o achieve CNR equal to that obtained in the scatter-free case. (C) 2001 Ame
rican Association of Physicists in Medicine.