The Physikalisch-Technische Bundesanstalt (PTB) has installed an X-ray radi
ometry beamline at the new electron storage ring BESSY II. The synchrotron
radiation is monochromatized by a four-crystal monochromator in the photon-
energy range from 1.75 keV to 10 keV. The known radiant power, determined w
ith a cryogenic electrical-substitution radiometer (ESR) as primary detecto
r standard, is used for the calibration of detectors such as semiconductor
photodiodes. Relative uncertainties of spectral responsivity well below 1%
have been achieved.