A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals

Citation
Si. Ziegler et al., A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals, EUR J NUCL, 28(2), 2001, pp. 136-143
Citations number
34
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
EUROPEAN JOURNAL OF NUCLEAR MEDICINE
ISSN journal
03406997 → ACNP
Volume
28
Issue
2
Year of publication
2001
Pages
136 - 143
Database
ISI
SICI code
0340-6997(200102)28:2<136:APHAPT>2.0.ZU;2-0
Abstract
To fully utilize positron emission tomography (PET) as a non-invasive tool for tissue characterization, dedicated instrumentation is being developed w hich is specially suited far imaging mice and rats. Semiconductor detectors , such as avalanche photodiodes (APDs), may offer an alternative to photomu ltiplier tubes for the readout of scintillation crystals. Since the scintil lation characteristics of lutetium oxyorthosilicate (LSO) are well matched to APDs, the combination of LSO and APDs seems favourable, and the goal of this study was to build a positron tomograph with LSO-APD modules to prove the feasibility of such an approach. A prototype PET scanner based on APD r eadout of small, individual LSO crystals was developed for tracer studies i n mice and rats. The tomograph consists of two sectors (86 mm distance), ea ch comprising three LSO-APD modules, which can be rotated for the acquisiti on of complete projections. In each module, small LSO crystals (3.7x3.7x 12 mm(3)) are individually coupled to one channel within matrices containing 2x8 square APDs (2.6x2.6 mm(2) sensitive area per channel). The list-made d ata are reconstructed with a penalized weighted least squares algorithm whi ch includes the spatially dependent line spread function of the tomograph. Basic performance parameters were measured with phantoms and first experime nts with rats and mice were conducted to introduce this methodology for bio medical imaging. The reconstructed field of view covers 68 mm, which is 80% of the total detector diameter. Image resolution was shown to be 2.4 mm wi thin the whole reconstructed field of view. Using a lower energy threshold of 450 keV, the system sensitivity was 350 Hz/MBq for a line source in air in the centre of the field of view. In a water-filled cylinder of 4.6 cm di ameter, the scatter fraction at the centre of the field of view was 16% (45 0 keV threshold). The count rate was linear up to 700 coincidence counts pe r second. In vivo studies of anaesthetized rats and mice showed the feasibi lity of in vivo imaging using this PET scanner. The first LSO-APD prototype tomograph has been successfully introduced for in vivo animal imaging. APD arrays in combination with LSO crystals offer new design possibilities for positron tomographs with finely granulated detector channels.