We report 4 eyes of a consecutive Series of 1299 that developed early decen
tration of a 10.5 mm diameter plate-haptic silicon intraocular lens (IOL) a
fter uneventful phacoemulsification. Ail eyes had an intact continuous curv
ilinear capsulorhexis (CCC) with the IOL placed in the capsular bag. After
an initial period of good vision, patients noted the onset of glare or mono
cular diplopia between 1 and 5 weeks after surgery. On examination, there w
as no significant anterior capsule contraction; however, the edge of the IO
L optic was visible in the undilated pupil. There was adhesion between the
anterior and posterior capsules at the margin of the CCC that maintained th
e IOL decentration, Decentration recurred in 1 eye after the IOL was rotate
d 90 degrees and recentered. Symptoms resolved in 3 eyes after the IOL was
removed and replaced With a rigid:IOL with a larger diameter optic. J Catar
act Refract Surg 2001; 27:330-3320 2001 ASCRS and ESCRS.