An automated analytical method for determination of polychlorinated bipheny
ls (PCBs) in fish was developed using supercritical fluid extraction (SFE)
in combination with an automated sample preparation instrument (Prep) and G
C/MS. By incorporating basic alumina with the sample in the extraction proc
ess, and optimizing the amount of carbon dioxide used, fish lipid was selec
tively reduced. The extract was cleaned up on a Florisil cartridge with Pre
p. The method was evaluated using naturally contaminated tissues and by com
parison of automated analytical method results with those obtained by the c
onventional method. Mean recovery of PCBs from 3 kinds of fish including ha
irtail, mackerel and yellowtail were 69.8%, 90.2% and 81.1%, respectively.
This method is less laborious and requires far less organic solvent than th
e conventional method, but produced comparable results.